Fabrication of Ultrasensitive Aluminum Oxide Nanoparticle Sensors for Potentiometric Determination of Ondansetron Dihydrate Hydrochloride in Pure and Pharmaceutical Forms

Authors

  • Dheaa Shamikh Zgear Higher Institute of Forensic Sciences, Al-Nahrain University, Jadiriya, Baghdad, Iraq.
  • Muataz Ali Higher Institute of Forensic Sciences, Al-Nahrain University, Jadiriya, Baghdad, Iraq. https://orcid.org/0000-0001-6825-9519
  • Amina M. Abaas Higher Institute of Forensic Sciences, Al-Nahrain University, Jadiriya, Baghdad, Iraq.
  • Rami A. Abdel-Rahem Department of Chemistry, Faculty of Arts and Sciences, University of Petra, Amman, Jordan.

DOI:

https://doi.org/10.22401/anjs.29.3.4212

Keywords:

Ion-selective membrane, Phosphomolybdic acid, Coated-wire electrode, Quality control.

Abstract

A highly precise potentiometric method for trace determination of Ondansetron dihydrate hydrochloride (OND) has been developed and established. A highly sensitive sensor of coated wire modified with aluminum oxide (Al2O3) was developed by incorporating trace amounts of (OND) with the ion-pairing reagent phosphomolybdic acid (PMA). Under the test conditions, the sensor exhibited a near-Nernstian response towards the drug over a linear range of 1.0×10-10 to 1.0×10-2 mol L-1 with limit of detection of 5.0x10-11 mol L-1 and slope of +59.30±0.05 mV decade-1 for (OND-PMA) coated wire sensor. The potentiometric sensor electrode showed stability up to 65 days and fast responsive time of 14 s. However, the developed potentiometric sensor for OND showed high selectivity to the mentioned drug in presence of various inorganic and some pharmaceutical species. The sensor response remained constant in a suitable potential at various pH from 2.5 to 9.0. Statistical analysis confirms that this potentiometric method is relatively very accurate and reproducible. The developed procedure was successfully applied to measure OND levels in both pure form and many commercial drug samples.

References

[1] Raposo, F.; Ibelli-Bianco, C. "Performance Parameters for Analytical Method Validation: Controversies and Discrepancies Among Numerous Guidelines". Trac. Trends Anal. Chem., 129: 115913, 2020. https://doi.org/10.1016/j.trac.2020.115913

[2] Kaur, D.; Kaur, J.; Kamal, S.S. "Development and Validation of a UV Spectrophotometric Method for Determination of Diacerein in Bulk and a Capsule Dosage Form". Indian J. Pharm. Sci. 81: 124-128, 2019. https://doi.org/10.4172/pharmaceutical-sciences.1000487

[3] Beena, P.; Krishnan, A.G.; Ullas, N.; Pillai, C.S.; Mathew, S.C.; Salim, S. "Development and Characterization of Orally Dissolving Tablet of a Poorly Soluble Antiemetic Drug". Res. J. Pharm. Technol. 14: 2172-2178, 2021. https://doi.org/10.52711/0974-360X.2021.00384

[4] Chanda, R.; Venkatesh, H.P.; Banerjee, J. "Formulation and Evaluation of Oral Disintegrating Tablets of Ondansetron Hydrochloride". Res. J. Pharm. Dosage Forms Technol. 11: 53-63, 2019. https://doi.org/10.5958/0975-4377.2019.00009.0

[5] Stoica, S.; Hogge, C.; Theeler, B.J. "Refractory Nausea and Vomiting Due to Central Nervous System Injury: A Focused Review". Life 15, 1021: 2025. https://doi.org/10.3390/life15071021

[6] British Pharmacopoeia Commission. British Pharmacopoeia 2013, Version 17. London: TSO (The Stationery Office), 2012.

[7] Keles, G.; Ataman, E.S.; Taskin, S.B.; Polatoglu, I.; Kurbanoglu, S. "Nanostructured Metal Oxide-Based Electrochemical Biosensors in Medical Diagnosis". Biosensors 14: 238, 2024. https://doi.org/10.3390/bios14050238

[8] Nagaraj, K. "Metal Oxides in Biomedicine: Advances in Imaging, Drug Delivery, Tissue Engineering, and Biosensing". Microchem. J. 215: 114481, 2025. https://doi.org/10.1016/j.microc.2025.114481

[9] Lenar, N.; Piech, R.; Wardak, C.; Paczosa-Bator, B. "Application of Metal Oxide Nanoparticles in the Field of Potentiometric Sensors: A Review". Membranes 13: 876, 2023. https://doi.org/10.3390/membranes1311876

[10] Ren, L.; Ji, M.; Zhuang, X.; Tian, C.; Shi, C.; Hao, X.W.; Luan, F.; Li, Y. "A Highly Stable Membrane-Free Potentiometric pH Sensor with Nickel Cobalt Sulfide as the Solid Contact Layer and Ion Recognition". Microchem. J. 201: 110679, 2024. https://doi.org/10.1016/j.microc.2024.110679

[11] Liao, C.; Zhong, L.; Tang, Y.; Sun, Z.; Lin, K.; Xu, L.; Lyu, Y.; He, D.; He, Y.; Ma, Y.; et al. "Solid-Contact Potentiometric Anion Sensing Based on Classic Silver/Silver Insoluble Salts Electrodes without Ion-Selective Membrane". Membranes 11: 959, 2021. https://doi.org/10.3390/membranes11120959

[12] Fathy, M.A.; Bühlmann, P. "Next-Generation Potentiometric Sensors: A Review of Flexible and Wearable Technologies". Biosensors 15, 51, 2025. https://doi.org/10.3390/bios15010051

[13] Huang, M.R.; Li, X.G.; Li, F.R. "Highly Sensing and Transducing Materials for Potentiometric Ion Sensors with Versatile Applicability". Prog. Mater. Sci. 125: 100885, 2022. https://doi.org/10.1016/j.pmatsci.2021.100885

[14] Ibrahim, M.; Ramadan, N.K.; Ibrahim, M.M.; Boltia, S.A. "Solid-Contact Ion-Selective Electrode for the Determination of Silver Ions Released from Silver Sulfadiazine in Sodium Hyaluronate Formulations". BMC Chem. 19: 167, 2025. https://doi.org/10.1186/s13065-025-01362-0

[15] Mariani, F.; Gualandi, I.; Schuhmann, W.; Scavetta, E. "Micro- and Nano-Devices for Electrochemical Sensing". Mikrochim. Acta 189: 459, 2022. https://doi.org/10.1007/s00604-022-05548-3

[16] Lingenfelter, P.; Bartoszewicz, B.; Migdalski, J.; Sokalski, T.; Bućko, M.M.; Filipek, R.; Lewenstam, A. "Reference Electrodes with Polymer-Based Membranes—Comprehensive Performance Characteristics". Membranes 9: 161, 2019. https://doi.org/10.3390/membranes9120161

[17] Dole, M. "The Theory of the Glass Electrode". J. Am. Chem. Soc. 53: 4260-4280, 1931. https://doi.org/10.1021/ja01363a003

[18] Rechnitz, G.A.; Hameka, H.F. "A Theory of Glass Electrode Response". Z. Anal. Chem., 214: 252-257, 1965. https://doi.org/10.1007/BF00512396

[19] Shirole, P.L.; Sanap, D.P.; Jadhav, K.R. "Development and Validation of a UV Spectrophotometric Method for Determination of Ondansetron Dihydrate Hydrochloride in Bulk and Tablet Dosage Form". Res. J. Pharm. Technol., 17: 1061-1064, 2024. https://doi.org/10.52711/0974-360X.2024.00166

[20] Jia, C.; Zhao, N.; Song, H. "Bioequivalence Analysis of Ondansetron Hydrochloride Tablets in Healthy Chinese Subjects: A Randomized, Open-Label, Two-Period Crossover Phase I Study". Drugs R D 24: 531-538, 2024. https://doi.org/10.1007/s40268-024-00493-3

[21] Shirole, P.; Sanap, D.P.; Jadhav, K. "Stability Indicating RP-HPLC Method Development for Ondansetron Hydrochloride Estimation in Bulk". Acta Pharm. Sci., 63: 134-151, 2025. https://doi.org/10.23893/1307-2080.APS6309

[22] Dyab, A.; Darwish, M.; Al Rahal, F.; Moustapha, A. "Selective Ion-Pair Potentiometric Sensor for the Determination of Ondansetron Dihydrate Hydrochloride in Different Matrixes". ChemistrySelect, 8; e202204551, 2023. https://doi.org/10.1002/slct.202204551

[23] Mohammed, A.A.; Khodair, Z.T.; Khadom, A.A. "Preparation, Characterization and Application of Al2O3 Nanoparticles for the Protection of Boiler Steel Tubes from High Temperature Corrosion". Ceram. Int., 46: 26945-26955, 2020. https://doi.org/10.1016/j.ceramint.2020.07.177

[24] Abass, A.M.; Abdoon, F.M. "Preparation of Cephalexin Monohydrate Sensors Based on the Synthesis of CeO2 NPs for Potentiometric Determination of Cephalexin Monohydrate in Pure and Commercial Products". Anal. Bioanal. Electrochem., 16: 943-956, 2024. https://doi.org/10.22034/abec.2024.717551

[25] Nassory, N.S.; Maki, S.A.; Ali, M.A. "Preparation and characterization of an atenolol selective electrode based on a PVC matrix membrane". Turk. J. Chem., 31: 75-82, 2007.

[26] Abbas, I.A.; Al-Mayalee, K.H. "Fabrication and Characterization of Al2O3 Nanoparticles Prepared via Pulse Laser Ablation Method". J. Kufa Chem. Sci., 5: 84-94, 2025. https://doi.org/10.36329/jkcm/2025/v5.i1.16878

[27] Toledo, R.R.; Santoyo, V.R.; Sánchez, C.D.M.; Rosales, M.M. "Effect of Aluminum Precursor on Physicochemical Properties of Al2O3 by Hydrolysis/Precipitation Method". Nova Scientia, 10: 83-99, 2018. https://doi.org/10.21640/ns.v10i20.1217

[28] Ali, M.; Alabdullah, S.S.M.; Al-Bassam, A.Z.M.; Fadhil, D.H. "New Electrochemical Sensors for Determination of Tamoxifen Based on Enhanced Polymer Nano Composite Deep Eutectic Solvent and Water Mixture as Ionophores". Anal. Bioanal. Electrochem., 16: 928-942, 2024. https://doi.org/10.22034/abec.2024.717550

[29] Gharbi, A.H.; Laouini, S.E.; Hemmami, H.; Bouafia, A.; Gherbi, M.T.; Ben Amor, I.; Hasan, G.G.; Abdullah, M.M.S.; Trzepieciński, T.; Abdullah, J.A.A. "Eco-Friendly Synthesis of Al2O3 Nanoparticles: Comprehensive Characterization Properties, Mechanics, and Photocatalytic Dye Adsorption Study". Coatings, 14: 848, 2024. https://doi.org/10.3390/coatings14070848

[30] Al-Zaidi, B.Y.S. "Analysis the Surface Morphology of the Porous Media by Using Atomic Force Microscope Technique". Al-Khwarizmi Eng. J., 13: 98-110, 2017. https://doi.org/10.22153/kej.2017.07.007

[31] Moaaz, E.M.; Fayed, A.S.; Abdel-Moety, E.M.; Rezk, M.R. "Innovative Sensors with Selectivity Enhancement by Molecularly Imprinted Polymers for the Concurrent Quantification of Donepezil and Memantine". RSC Adv., 15: 18475-18489, 2025. https://doi.org/10.1039/d5ra02850g

[32] Shahzad, U.; Saeed, M.; Marwani, H.M.; Al-Humaidi, J.Y.; Rehman, S.U.; Althomali, R.H.; Awual, M.R.; Rahman, M.M. "Recent Progress on Potentiometric Sensor Applications Based on Nanoscale Metal Oxides: A Comprehensive Review." Crit. Rev. Anal. Chem. 55, 1081–1098, 2024. https://doi.org/10.1080/10408347.2024.2337876.

[33] Mirabootalebi, S.O.; Liu, Y. "Recent advances in nanomaterial-based solid-contact ion-selective electrodes." Analyst 149, 3694–3710, 2024. https://doi.org/10.1039/d4an00334a.

Downloads

Published

2026-09-15

How to Cite

(1)
Shamikh Zgear, D. .; Ali, M.; M. Abaas, A. .; A. Abdel-Rahem, R. . Fabrication of Ultrasensitive Aluminum Oxide Nanoparticle Sensors for Potentiometric Determination of Ondansetron Dihydrate Hydrochloride in Pure and Pharmaceutical Forms. Al-Nahrain J. Sci. 2026, 29 (3), 26-36. https://doi.org/10.22401/anjs.29.3.4212.