RADON DOSIMETRY USING CR_39 AS A TRACK DETECTOR

Authors

  • Nada F College of Science, Al-Nahrain University.
  • Nahida J Hameed Applied science Department, University of Technology.

Keywords:

NON

Abstract

Solid state nuclear detector (CR-39) was used to obtain the concentration level of radon and itsdaughters, by using (Ra-266) of (10.54mCi) activity. It was employed for radon concentration andpotential alpha energy exposure level. It is found increasing in track density with (absorbed dose,WLM, radon concentration in (PCi/liter), and potential alpha energy (PAEC)). The calibrationcurves were drawn by taking the linear part from the above curves. There is decreasing inabsorbed dose with distance from the source. The linear response determine the detectorrange of (300-3X104 Bq/m3), and the calibration factors was obtained for calibration(0.2 } 0.085 track/ mm2 per Bq/m3).

Published

2010-06-01

Issue

Section

Articles

How to Cite

[1]
“RADON DOSIMETRY USING CR_39 AS A TRACK DETECTOR”, ANJS, vol. 13, no. 2, pp. 124–128, Jun. 2010, Accessed: Apr. 23, 2024. [Online]. Available: https://anjs.edu.iq/index.php/anjs/article/view/1104