DESIGN AND CONSTRUCTION OF A DIAGNOSTIC PULSED X-RAY TUBE

Authors

  • Mohamed Ibrahim Department of Chemical Engineering, University of Technology.

Keywords:

Pulsed x ray, cold cathode, Marx generatoin

Abstract

A pulsed X-ray Tube based on the field emission cold cathode principle has been designed andconstructed. The device produced X- ray pulses of duration of (1.2) μ sec and intensities of (90) mRat a distance of (10) cm from the window, when a (150) KV is discharged across the electrodes ofthe tube. The intensity of the x ray radiation at various distances was determined at various cathodeanode spacing to determine its optimum value. Special efforts were made in the design to bedemountable, easy to maintain, and offering a practical possibility for future industrial and medicalapplication

 

Published

2018-07-31

Issue

Section

Articles

How to Cite

[1]
“DESIGN AND CONSTRUCTION OF A DIAGNOSTIC PULSED X-RAY TUBE”, ANJS, vol. 13, no. 3, pp. 51–57, Jul. 2018, Accessed: Apr. 26, 2024. [Online]. Available: https://anjs.edu.iq/index.php/anjs/article/view/1038